Model 2241-2
Digital Scaler/Ratemeter
Click here to return
to INSTRUMENT SALES
INDICATED USE: General purpose survey,
gross counting
COMPATIBLE DETECTORS: G-M, proportional,
scintillation
CONNECTOR: Series "C" (others available
on request)
AUDIO: Built in unimorph speaker
with ON/OFF switch (greater than 60 dB at 2 feet)
ALERT/ALARM: Indicated by annunciator
on display and audible tone
DISPLAY: 4 digit LCD display with
0.5"(1.3cm) high digits, separate annunciators for display units, alert,
alarm, low battery, detector overload, counting overflow, and scaler counting
BACKLIGHT: Push-button to activate
RATEMETER: Can display in R/hr,
Sv/hr, cpm, or cps when control switch is in RATEMETER position
DISPLAY RANGE: Auto ranging from
0.0 R/hr - 9999 R/hr; 0.000 Sv/hr - 9999 Sv/hr; 0 cpm - 999k cpm; or 0
cps - 100k cps
LINEARITY: Reading within plus
or minus 10% of true value with detector connected
SCALER: Activated by push-button
in handle (count time adjustable from 1 - 9999 seconds in 1 second intervals)
CALIBRATION CONTROLS: Accessible
from front of instrument (protective cover provided)
HIGH VOLTAGE: Adjustable from 200
- 2500 volts
DISCRIMINATOR: Adjustable from
2 - 100 mV
OVERLOAD: Indicated by OVERLOAD
on display (adjustable depending on detector selected)
RESET: Push-button to zero display,
acknowledge and/or reset alarm
POWER: 2 each "D" cell batteries
(housed in sealed compartment that is externally accessible)
BATTERY LIFE: Typically 200 hours
with alkaline batteries (low battery indicated on display)
CONSTRUCTION: Cast and drawn aluminum
with beige polyurethane enamel paint
TEMPERATURE RANGE: -4 degrees F(-20
degrees C) to 122 degrees F(50 degrees C)
May be certified for operation from -40
degrees F(-40 degrees C)to 150 degrees F(65 degrees C)
SIZE: 6.5"(16.5cm)H X 3.5"(8.9cm)W
X 8.5"(21.6cm)L
WEIGHT: 3.5 lbs (1.6kg) including
batteries
The following parameters are adjusted by
removable switchboard. Once initial setup and calibration have been completed
the switchboard can be removed without affecting normal operation of the
instrument (Model 2241 may be ordered without switchboard)
BACKLIGHT "ON" TIME: 5, 30, 60,
90, 180, or 240 seconds for the backlight to stay "ON" when activated.
SET MINIMUM DISPLAY: Allows lower
limit of auto-ranging display to be fixed. For example, display can be
set to not show values less than 1 R/hr.
RS-232 DATA DUMP MODE: Enables
or disables dump mode to RS-232 port. When enabled data will be dumped
every 2 seconds
RS-232 DETECTOR SETUP MODE: Allows
for detector parameters to be input by RS- 232 port
BAUD RATE: 150, 300, 600, 1200,
2400, 4800, 9600, or 19200 bps.
DEAD TIME: Adjustable from 0 -
9999 seconds
CALIBRATION CONSTANT: Adjustable
from 0.001 - 280 X 109 counts/display unit
DISPLAY UNITS: Can read in R/hr,
Sv/hr, cpm, or cps
TIME BASE: Can read in seconds
or minutes
AUDIO DIVIDE: 1, 10, 100, or 1000
events-per-click
RESPONSE:
FIXED: FAST response time adjustable
from approximately 2 - 50 seconds. SLOW response is 5 times fast setting
VARIABLE: Dependant on number of
counts present. Typical times FAST - 4 - 25 seconds, SLOW - 4-60 seconds
(All times stated from 10% to 90% of final reading)
RATEMETER ALERT/ALARM: Set at any
point corresponding to pre-selected ratemeter range
SCALER ALARM: Adjustable from 1
- 999999 counts
SCALER COUNT TIME: Adjustable from
1 - 9999 seconds

|