Model 3-97 Gamma Scintillation Survey meter with Model 44-38 G-M Detector
Click here to return
to INSTRUMENT SALES
INDICATED USE: Beta-gamma survey
DETECTORS: Controlled by selector
switch
INTERNAL: 1" X 1" NaI(Tl) scintillator
SENSITIVITY: Typically 175 cpm/microR/hr
( Cs-137 gamma)
EXTERNAL: Model 44-38 energy compensated
G-M detector (others available)
SENSITIVITY: Typically 1200 cpm/mR/hr
(Cs-137 gamma)
CONNECTOR: Series "C" (others
available)
METER DIAL: 0 - 2 mR/hr, 0 - 2.4k
cpm, BAT TEST (others available )
MULTIPLIERS: X0.1, X1, X10, X100
LINEARITY: Reading within plus
or minus 10% of true value with detector connected
AUDIO: Built in unimorph speaker
with ON/OFF switch (greater than 60 dB at 2 feet )
CALIBRATION CONTROLS: Accessible
from front of instrument (protective cover provided )
HIGH VOLTAGE: Adjustable from 200
- 1500 volts
THRESHOLD: 30 mV plus or minus
10 mV
RESPONSE: Toggle switch for FAST
(4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
RESET: Push-button to zero meter
POWER: 2 each "D" cell batteries
(housed in sealed compartment that is externally accessible )
BATTERY LIFE: Typically 600 hours
with alkaline batteries (battery condition can be checked on meter
)
METER: 2.5" (6.4 cm) arc, 1 mA
analog type
CONSTRUCTION: Cast and drawn aluminum
with beige polyurethane enamel paint
TEMPERATURE RANGE: -4 degrees F(-20
degrees C) to 122 degrees F(50 degrees C)
May be certified for operation from -40
degrees F(-40 degrees C) to 150 degrees F(65 degrees C)
SIZE: 9.5"(24.1cm)H X 3.5"(8.9cm)W
X 8.5"(21.6cm)L
WEIGHT: 4.4 lbs (2 kg) including
batteries and detector

|